| Thromboelastograph | ||
| Order Code: TEG
Epic Lab Code: LAB5831 Order Form: A-1a Miscellaneous Request or Epic Req |
Hemostasis/Thrombosis 6240 RCP 356-3573 |
|
Specimen: |
Whole Blood | ||
Collection Medium: |
| ||
Minimum: |
Full draw of 2.7 mL Light Blue Sodium Citrate tube | ||
Delivery Instructions: |
Deliver to laboratory immediately after collection. | ||
Testing Schedule: |
24 hrs/day, 7 days a week, including holidays. | ||
Analytic Time: |
2 hours (upon receipt in laboratory) | ||
Reference Range: |
R: 2 - 8 minutes K: 1 - 3 minutes Angle: 55 - 78 degrees MA: 51 - 69 mm LY30: 0 - 8 % | ||
Comments: |
This test needs a separate tube! It CANNOT be performed with any other coagulation tests. | ||
Methodology: |
A stationary pin attached to a torsion wire is immersed into whole blood. When the first fibrin forms, it begins to bind the cup and pin, causing the pin to oscillate in phase with the clot. A mechanical-electrical transducer converts the rotation movement of the pin to an electrical signal which is monitored by a computer. | ||
CPT Code: |
85347, 85384 (x2), 85390, 85576 |